RF Output Control Commands [SOURce[1]:]POWer:ATTenuation:AUTO Disadvantages When the step attenuator is locked, the output power dynamic range is limited to the vernier range at the current output frequency. The vernier range extends from a lower limit that is typically 5 dB lower than the specied value for that range to an upper limit that is frequency dependent on the synthesizer output frequency. NOTE In external diode detector leveling or external power meter leveling mode, the attenuator is always locked in the current range and can not be unlocked using this function. Query Syntax 2 2 3 3 SOURce 1 : POWer:ATTenuation:AUTO? Returned format: state<NL> Where: state ::= \+1" if the attenuator hold function is currently o or \+0" if the attenuator hold function is currently on or set to \once". 2j-8