Download Agilent Technologies PSG Programming instructions

Transcript
Command Reference
Power Subsystem ([:SOURce])
:POWer:ALC:SOURce
Supported
All
[:SOURce]:POWer:ALC:SOURce INTernal|DIODe|MMHead
[:SOURce]:POWer:ALC:SOURce?
This command enables you to select the ALC leveling source.
*RST
INT
Choices
INTernal
Key Entry
Internal
Remarks
N/A
DIODe
MMHead
Ext Detector
Source Module
:POWer:ALC:SOURce:EXTernal:COUPling
Supported
All
[:SOURce]:POWer:ALC:SOURce:EXTernal:COUPling <value>DB
[:SOURce]:POWer:ALC:SOURce:EXTernal:COUPling?
This command sets the external detector coupling factor.
*RST
+1.60000000E+001
Range
−200 to 200
Key Entry
Ext Detector Coupling Factor
Remarks
Use this command when DIODe is the selected ALC leveling source.
Refer to “:POWer:ALC:SOURce” for the source selection.
Chapter 4