Transducer Care 10 Disinfectants Compatibility Table The table that follows lists the disinfectants compatible with the transducers available for your system. Other low- and intermediate-level disinfectants marketed for use on medical instruments and based on quaternary ammonium compounds (QUATS) or sodium hypochlorite (NaOCl), equal to or less than 0.6%, are approved for use. NOTE A 10% bleach solution typically provides a solution that is less than 0.6% NaOCl. Additionally, products that contain 70% or less isopropyl alcohol (IPA) are acceptable for use on the transducer portion of the device only. (Do not use alcohol-based products on transducer cables and strain reliefs.) Those disinfectant types must be used only in a spray or wipe application. CAUTION The preceding recommendations do not apply to TEE transducers. For more information: • Philips Ultrasound Transducer Care Web site (http://www.medical.philips.com/main/products/ultrasound/transducers/transducercare/) • In North America, call Philips Ultrasound Customer Service at 800-722-9377. • Outside North America, contact your local Philips Ultrasound representative. CX50 User Manual 4535 613 06532 231