Download CONTENTS
Transcript
System Information Messages Background messages Indicator b B F Message RBC background high WBC background high Differential background high Baso background high F H p HGB background high PLT background high Description Last accepted background result: RBC ≥ 0.01 x 1012 cells/L Last accepted background result: WBC ≥ 0.1 x 109 cells/L More than 100 cells detected during the 4-Diff background procedure. More than 100 cells detected during the Baso background procedure. Last accepted background result: HGB ≥ 0.2 g/dL Last accepted background result: PLT ≥ 10 x 109 cells/L Action · Perform background measurement · Check cleanliness of the reagents and the Quintus system · If background, is not acceptable perform [CLEAN] · Repeat background measurement · If issue persists see Section 9.3: High background cleaning Treat result as “reduced reliability” Distribution Indicators (RBC,WBC) Indicator Message S WBC Measurement Statistics Warning s RBC Measurement Statistics Warning Description The distribution of the WBC detection changes in time. It points to clogging, nonhomogeneous sample, cold sample, partial coagulation of the sample. The distribution of the RBC detection changes in time. It points to clogging, nonhomogeneous sample, cold sample, partial coagulation of the sample. Action · Re-analyze sample · Check the homogeneity, the temperature and the coagulation of the sample · Verify system has been correctly maintained, see Section 9.1 - 9.4 Clogging/Leakage Indicator (WBC, RBC) Indicator Message C WBC clogging c RBC clogging V WBC Vacuum Warning V RBC Vacuum Warning Indicator Message Description Drift in the probe voltage of the WBC capillary. The main reason for this drift is the clogging of the capillary. Drift in the probe voltage of the RBC capillary. The main reason for this drift is the clogging of the capillary. There is a (partial) clogging or leakage in the WBC part of measurement system. A faulty/ worn-out pump can create vacuum errors as well. There is a (partial) clogging or leakage in the RBC part of measurement system. A faulty/ worn-out pump can create vacuum errors as well. Action · Re-analyze sample · Verify system has been correctly maintained, see Section 9.1 - 9.4 · Re-analyze sample · Perform a self-test · Verify system has been correctly maintained, see Section 9.1 - 9.4 · If vacuum problems persist then contact service. Linearity Indicator (RBC) M Out of RBC linearity range m Inconsistency in RBC cell concentration Description Inconsistency in RBC cell concentration. 80 Action · Check the sample homogeneity. · Re-analyze sample · Repeat the sample with manual pre-dilution. · Check the sample homogeneity. · Re-analyze sample