Download CONTENTS
Transcript
Step Action Note The Standby time is activated after a background, control, or sample analysis. Screensaver time: Use scroll buttons to change the how long the analyzer should wait before the screensaver function is activated. · Choose between 3, 5, 10, 15, 30, and 60 minutes · Recommended setup time is 15 minutes To activate the display simply touch the screen. Overnight Clean: Use scroll buttons to change how long the analyzer should wait before running the overnight cleaning. · When this function is activated, the software will perform a minor fluidic operation to make sure that there are fresh reagents in measurement critical sections of the tubing system. · Choose between 0, 4, 8, 12, and 24 hours · Recommended setup time is 8 hours 0 hour selection means no cleaning cycle will be performed 4 Note 5 Note 4.10 Profile Limits Settings Profile Limits Setup Indicative normal ranges (profile limits) are provided in this instrument. It is recommended to establish local reference ranges (normal ranges) for your laboratory. (See CLSI standard C28-A2 for guidance on how to establish these ranges and examples of normal ranges in the reference documents listed at the end of this section.) Step 1 2 Action Start by pressing [ADVANCED] from the MENU tab. Press [SETUP] and then [PROFILE LIMITS] to enter setup menu. Figure 4.12 To change the low or high profile limits, select the limits box and type in new limit. Normal Range References 1. Cheng C, Chan J, Cembrowski G, van Assendelft O. Complete Blood Count Reference Interval Diagrams Derived from NHANES III: Stratification by Age, Sex, and Race Laboratory Hematology 10:42-53 2. Nordin G, et al. A multicentre study of reference intervals for haemoglobin, basic blood cell counts and erythrocyte indices in the adult population of the Nordic countries Scand J Clin Lab Invest 2004; 64: 385-398 3. How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline – Second Edition. CLSI C28-A2 33