Download Scale Tab
Transcript
RF Measurements Freq & RBW Tab for ACPR and MCPR Displays Freq & RBW Tab for ACPR and MCPR Displays The Freq & RBW tab specifies frequency parameters for the Channel Power & ACPR measurements and MCPR measurement. Setting Description Meas Freq Specifies the center/measurement frequency. Step The Step control sets the increment/decrement size for the adjustment of the center frequency. If Auto is enabled, the analyzer will adjust the Step size as required. RBW Select Auto or Manual. Adjusts the RBW for the entire measurement. This setting is Independent of the Spectrum view's RBW setting. VBW Adjusts the VBW (Video Bandwidth) value. VBW Maximum = current RBW value. VBW Minimum = 1/10,000 RBW setting. xxx NOTE. While the center frequency and step settings on this control tab are shared between ACPR, MCPR and Occupied Bandwidth, the RBW setting is not shared. The RBW setting is unique for each measurement. VBW The VBW setting enables/disables the Video Bandwidth filter. VBW is used in traditional swept analyzers to reduce the effect of noise on the displayed signal. The VBW algorithm in the analyzer emulates the VBW filters of traditional swept analyzers. The maximum VBW value is the current RBW setting. The minimum VBW value is 1/10,000 of the RBW setting. Measurement Params for ACPR and MCPR Displays The Measurement Params tab is where you set parameters that control the ACPR and MCPR measurements. SignalVu-PC Printable Help 117