Chapter 1, Calibrating the Test Set Zeroing Average Power Measurements Zeroing Average Power Measurements Approximate length of time: 2 seconds Average Power measurements should be zeroed before each measurement or series of measurements. NOTE: A misleading Average Power measurement may appear when low (or no) signal power is applied to the RF Input! When the RF generator’s output port selection is RF IN/ OUT, some of the signal energy from the Test Set’s generator is detected by the Test Set’s broadband average power meter. This condition does not affect typical CDMA measurements for two reasons: 1) During Average Power measurements CDMA generator levels are too low to introduce significant energy to the power detector. 2) When the generator level is high enough to introduce significant energy to the power detector, the mobile station’s signal power should be within the range of Channel Power measurements. Channel power measurements are frequency-selective, and do not detect significant energy from the Test Set’s generator, which is tuned 45 MHz away from the analyzer. 36 N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcal.fb