Chapter 4, CDMA Transmitter Tests Measuring the Range of Open Loop Output Power Measuring the Range of Open Loop Output Power The Test Set measures the range over which the mobile-station-under-test (MSUT) can adjust its Effective Radiated Power (ERP) in response to the power level it receives from the Test Set. This test is performed according to TIA IS-98 standards, except that Îor is set to −35 dBm/BW instead of the recommended value of −25 dBm/BW for the first test. Test Prerequisites • The Channel Power measurement should be calibrated at least as often as the following conditions arise: • 5 °C change in operating temperature • Power cycle • Daily See "Calibrating Channel Power Measurements" on page 29 if necessary, and then return to this procedure. NOTE: The following procedure assumes that an instrument PRESET will be performed as part of setting up a call. This clears all settings from other tests that may affect the accuracy of this test, and allows the procedure to begin from a known instrument state. N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb Chapter 4 CDMA Transmitter Tests 175