Chapter 4, CDMA Transmitter Tests
Measuring the Range of Open Loop Output Power
Measuring the Range of Open Loop Output Power
The Test Set measures the range over which the mobile-station-under-test
(MSUT) can adjust its Effective Radiated Power (ERP) in response to the power
level it receives from the Test Set. This test is performed according to TIA IS-98
standards, except that Îor is set to −35 dBm/BW instead of the recommended
value of −25 dBm/BW for the first test.
Test Prerequisites
•
The Channel Power measurement should be calibrated at least as often as the following
conditions arise:
• 5 °C change in operating temperature
• Power cycle
• Daily
See "Calibrating Channel Power Measurements" on page 29 if necessary, and
then return to this procedure.
NOTE:
The following procedure assumes that an instrument PRESET will be performed as part of
setting up a call. This clears all settings from other tests that may affect the accuracy of this
test, and allows the procedure to begin from a known instrument state.
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb
Chapter 4
CDMA Transmitter Tests