Maximum Heat Dissipation of X-ray Head Assembly Leakage Radiation Minimum mAs Weight of X-ray Head 2 kJ/min. max. 1.0mGy/h at 1m Panoramic 6.0 mAs Cephalometric 3.5 mAs 13.8Kg (with collimators) Auto Exposure (Not applied for Cephalometric) Setting parameters +4 - -4 (9 steps) Digital Display Maximum possible excursion 60kV – 80kV 1mA – 10mA Minimum exposure time 4.3 seconds Verification method by test piece Power requirements Input Voltage Fuse at the distribution panel Power Consumption Rated input amperes Power line resistance Mechanical parameters Panoramic Cephalometric Magnification X-ray Field EX-2: AC220/230/240V, 50/60 Hz single phase EX-2:16A, 220/230/240V, slow (220/230/240V) EX-2: max. 2.0KVA EX-2: max. 9A/ 8A/ 8A (AC220/230/240V, with operation) EX-2: max. 1.0A (220/230/240V, stand by) EX-2: max. 1 Ohm (220/230/240V) SID 520mm SID 1,650mm Standard Panoramic: 1.3×,1.6× Maxillary Sinus Panoramic: 1.5× Pedodontic Panoramic: 1.3×, 1.6× Shadow Reduction Panoramic: 1.3×, 1.6× Orthoradial Panoramic: 1.3×, 1.6× Scanograms: 1.3×, Cephalometric: 1.1× 6 × 146 mm (+0/-30%): Panoramic 6 × 220 mm (+0/-30%): Cephalometric Weight Panoramic Approximately 184 kg Boxed Cephalometric Approximately 268 kg Boxed Outer Dimensions Main Unit W 1,020 × D 1,330 × H 2,355mm (H 2,185mm Option) (Panoramic, Upgradable Cephalometric) W 2,000 × D 1,330 × H 2,355mm (H 2,185mm Option) (with Cephalometric) Control box W 70 × D 40 × H 115mm Vertical Height of Focal Spot 1,055 ~ 1,775mm (Panoramic) 970 ~ 1,605mm (Option) 1,125 ~ 1,775mm (with Cephalometric, Upgradable Cephalometric) 1,040 ~ 1,605mm (Option) Patient Positioning Auto focus using light sensor for distance measurement and electrically operated positioning system Patient Positioning Beam Class II Laser. Accordance with 21CFR Part1040.10 and IEC60825-1. Attenuation equivalent of Panoramic less than 1.7 mmAl Patient head stabilizer Cephalometric less than 1.7 mmAl 89 Operation 2007-09-19