Download K. Fowler (CRC, 2008) BBS - Web server administration

Transcript
Programmer for Implanted Stimulators
319
TABLE 13.4
Example Portion from a Survey for the Usability of the Programmer
User Test Survey—Summary Example
Patient Tutorial Questions
Yes
No
Will these instructions be clear to your patient population?
Will pressing the buttons be easy for your patients?
Do you think the function of the OFF button will be obvious to
your patients?
Will drawing on the body maps be easy for your patients?
Do you think the use of the front and back body maps will be
obvious to your patients?
Do you think the Right and Left labels on the body maps will be
effective with your patients?
Will marking the rating scales be easy for your patients?
Do you think your patients will be able to adjust the stimulation
amplitude easily?
Do you think showing this tutorial to your patients will improve
their comfort level with neurostimulation?
7
8
4
1
0
4
7
4
1
4
6
2
8
8
0
0
8
0
Average
6.7
1.3
Std. Dev.
1.7
1.7
83%
17%
% Agreement
Comments
Have a way to stop and then resume the tutorial
Voice was pleasant
Several commented on how well worded it was
One physician thought graphics were ‘‘too Mickey Mouse’’
One physician wants branded styluses—one with Alabama, the other with Auburn
Patient Tab Questions
Yes
No
Does the Patient Tab provide the basic information you
need?
Do the buttons on the right side of the Patient Tab make
sense to you?
Is the Leads information on the Patient Tab clear to you?
Does the automated Read Implant function make sense to
you?
Are you likely to use the Remarks section?
6
1
8
0
8
8
0
0
8
0
Average
7.6
0.2
Std. Dev.
0.9
% Agreement
97%
0.4
3%
Comments
Undo Changes button was not intuitive
Training is a concern. If staff is to use programmer then they must be trained
Consider have more information collection: male/female selection, age, diagnosis field
(could use Remarks section) with boxes to check like radicularopathy, CRPS, spinal
stenosis, and so forth, suggested CPT codes
Read Implant button is not intuitive. Use ‘‘Read Serial Number’’ instead or consider
‘‘Read Generator’’
Use ‘‘Exit’’ instead of ‘‘LogOff’’ for button
(Continued)