Using Remote Control
Use the SCPI/IEEE-488 Interface
Use the SCPI/IEEE-488 Interface
After you have connected the 156MTS and SCPI controller, and
configured the test set’s interface parameters, you are ready to activate
SCPI mode.
1. Using the SCPI/IEEE-488 interface on the controller device, send
the reset string “*RST” to the test set, or use the :SYST:REM
command.
These commands puts the test set in SCPI/HP-IB mode. *RST sets
all parameters to their default values; :SYST:REM maintains the
current parameter settings. Refer to the SCPI/HP-IB Programmer’s
Manual for more information.
2. When the test set is in SCPI mode, the display changes to show the
following:
1 SCPI Remote Mode Final: 00:00:00.00
Trouble Scan
>>>>>>>
SONET LOS:
HIS
DS3 LOS:
HIS
SONET LOF
HIS
DS3 LOP:
ON
SONET LOPNTR: ON
DS3 FFM:
ON
PATH CV: 18273645
More...
Enable Display of Results and Alarms
SONET>Off
ATM>Off
DS3>Off
PLCP>Off
T-Carrier>Off
VT1.5>Off
Results Level>Detail
Trouble Scan>Off
Note: When the test set is in SCPI mode, most front-panel controls are inoperative.
Setup operations must be done using SCPI commands from the controller.
3. Use the FIELD and VALUE keys to configure which results are
shown on the test set display while the unit is in SCPI mode.
4. Next use FIELD to select Results Level and then use VALUE to set
the results level for the level of measurement detail you want.
General
5. Use the RESULT keys to scroll through measurement screens in the
upper half of the display. The measurement screens available
depend on the results enabled in the lower half of the display.