Chromira ProLab Troubleshooting Drum Tension Spring Reference If there was some other problem, you may see in the Diagnostics. This white circle-indicators show that the Drum Motor was Not Tested. This indicates that something prevented the Drum Tensioning Motor and Spring from referencing properly. Check the following: - The Drum’s Spring Cage is secure to the end of the Drum. Check this by simply grasping the Drum with one hand, and the outer plate of the Spring Cage with the other. Try to move these in opposite directions – they should feel connected as one solid unit. If not, screw in the Spring Cage, which has reverse threads, and re-try this Test. - Check that the Drum’s movement is free. You will feel the gentle force of the Spring, but the movement should be smooth, free, and quiet. If not, check for obstructions. ProLab Troubleshooting Guide Copyright 2010 Page 21 of 75