TÜV Rheinland Energie und Umwelt GmbH
Air Pollution Control
Page 92 of 424
6.1
Report on the suitability test of the ambient air quality measuring system
Model 5030i SHARP with PM2.5 pre-separator of the company Thermo
Fisher Scientific for the component suspended particulate matter PM2.5,
Report No.: 936/21209885/F
5.2.5 Operating states
The measuring system shall allow the control of important operating states by telemetrically transmitted status signals.
6.2
Equipment
PC for data acquisition.
6.3
Performance of test
A PC was connected locally via Ethernet to the AMS and the transfer of data including system status was tested.
Furthermore, it is possible to monitor and operate the instrument via serial interface.
It is also easy to monitor and operate the AMS remotely with an appropriate router or modem.
6.4
Evaluation
The measuring system enables an extensive telemetric control and operation in a variety of
ways (Ethernet, RS232, RS485).
6.5
Assessment
The measuring systems can be extensively monitored and operated from an external PC via
modem or router.
Minimum requirement fulfilled? yes
6.6
Detailed presentation of test results
Not required for this test.
740300_2013_936_21209885F_EN.docx