MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Conductivity - Diagnostics Menu - Calibration Data 6.14
1000uS 25.0C
RUN MODE
The Calibration Data menu is a series of read only screens which
allow the user to view the data collected during the last calibration.
DOWN
CALIBRATION
DOWN
UTILITIES
DOWN
SETUP
DOWN
DIAGNOSTICS
FIRMWARE REV.
DOWN
Press
to view the first
Calibration Data screen.
CALIBRATION DATA
TYPICAL DISPLAYS
FOR MANUAL CAL
CALIBRATION DATA
MODE 1PT
>
Press DOWN to view the next
Calibration Data screen.
DOWN
CALIBRATION DATA
1P 1000/1000 Ω
Calibration Mode
1 point calibration
>
DOWN
>
Calculated Cell Constant
CALIBRATION DATA
K FACTOR 1.0000
Calibration Mode
Dry Cal
Press DOWN to view the next
Calibration Data screen.
DOWN
CALIBRATION DATA
1P - - - -
>
No Buffer used in
this Calibration Mode
DOWN
>
DOWN
CALIBRATION DATA
CAL TEMP 25.2C
CALIBRATION DATA
MODE DRY
Conductivity of Buffer/
Input Resistance
DOWN
CALIBRATION DATA
K FACTOR 0.9774
TYPCIAL DISPLAYS
FOR DRY CAL
Cell Constant Setting
>
DOWN
>
Probe Temperature at which
calibration was performed =
25.2° C
CALIBRATION DATA
CAL TEMP - - - DOWN
>
Dry Cal does not use a buffer, so
Probe Temperature was
not recorded
NOTE
PRESS THE UP AND DOWN KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
Conductivity - Diagnostics Menu - Calibration Data 6.14
Page 60