Download EX1629 - VTI Instruments
Transcript
www.vtiinstruments.com MSW (MAX Limit exceeded status) 31 LSW (MIN Limit exceeded status) 15 0 = Conf. data is less than the MAX for Conf. Source 0 1 = Conf. data is greater than the MAX for Conf. Source 0 0 0 = Conf. data is greater than the MIN for Conf. Source 0 1 = Conf. data is less than the MIN for Conf. Source 0 Confidence Limit Check Detailed Result (Available per Bridge channel) NOTE The source-to-bit mapping is constant, regardless of confidence scanlist configuration. For example, whether or not sources 0 and 1 are enabled in the confidence scanlist, for instance, source 2‟s MIN Limit Exceeded Bit and MAX Limit Exceeded Bit are always bits 2 and 18, respectively. The confidence source mapping follows the same ordering as the source # define in vtex1629.h i.e. if sources 3, 8, and 10 are selected then they are reported in that order. Confidence sources that are not part of the confidence scanlist are not reported and will have their bit-fields set to 0. Confidence values are reported at a maximum frequency of 500 Hz. This mode is supported up to 1 kHz sampling rate. Hence, at 1 kHz, every other packet will contain confidence information. The datapage size is 248 words when it has full confidence information i.e. confidence data and full limit check values, and is 24 words when it has no confidence information. Hence, the total data rate = ((248+24)/2)*4*8*1000 samples/second= 4.352 Mb/s. EXAMPLE ViSession instrumentHandle; ViStatus status; ViInt32 channel = 10; ViReal64 min = 0; ViReal64 max = 0; ViInt32 confSrcEnum = 0; … status = vtex1629_get_confidence_limit (instrumentHandle, channel, confSrcEnum, &min, &max); EX1629 Command Set 143