VTI Instruments Corp. vtex1629_get_strain_units FUNCTION PROTOTYPE ViStatus vtex1629_get_strain_units (ViSession vi, ViInt32 channel, ViPBoolean microStrain); FUNCTION PARAMETERS vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the session is closed. channel = an integer input value that specifies the channel for which the strain units will be returned. Valid input values: 0 to 47. microStrain = a Boolean return value that indicates whether the EX1629 will return strain measurements in units of strain (ε) or microstrain (µε). A value of “1” indicates that the EX1629 is configured to return microstrain units for the given channel, whereas a value of “0” indicates the EX1629 is configured to return strain units for the given channel. DATA ITEM RESET VALUE Not applicable to this function. DESCRIPTION This function queries and returns the configured strain units for a given channel. Each channel can be configured to return strain measurements in strain or microstrain (1 strain (ε) = 1x10 6 microstrain (µε)). This setting has no effect for non-strain EU conversions. EXAMPLE ViSession instrumentHandle; ViStatus status; ViBoolean units_are_microstrain; … … status = vtex1629_get_strain_units(instrumentHandle, 0, &units_are_microstrain); 190 EX1629 Command Set