Download Model 42i-LS - User Manual
Transcript
Servicing NO2-to-NO Converter Replacement reaction chamber and the quartz window. To clean these surfaces use cotton swabs and methanol. 6. To continue removing rear of reaction chamber remove the three socket head screws holding it to cooler, being careful to keep quartz window and red filter in cooler body. 7. To reinstall reaction chamber, follow previous steps in reverse, making sure to backfill the cooler with dry air or nitrogen prior to installing reaction chamber. 8. Re-install the measurement bench. Refer to “Removing the Measurement Bench and Lowering the Partition Panel” in this chapter. NO2-to-NO Converter Replacement Use the following procedure to replace the converter (Figure 7-12). Equipment Required: NO2-to-NO Converter Wrench, 7/16-inch Wrench, 9/16-inch Wrench, 1/2-inch Wrench, 5/8-inch Screwdriver Nut driver, 1/4-inch Nut driver, 5/16-inch CAUTION Avoid contact with converter heated components. Allow converter to cool to room temperature before handling converter components. Thermo Fisher Scientific Model 42i Low Source Instruction Manual 7-21