Download user Manual
Transcript
3.12 To achieve a good calibration some basic rules must be observed: 1. The Oxygen sensors must be exposed to the pure calibrating gas. So for an air calibration, the sensors must be flushed with air. Just taking the hoses off is NOT sufficient, as pockets of gas can be enveloping the sensors. 2. DO NOT execute calibrations to remedy a rebreather whose PPO2 accuracy is drifting over short periods of time. Some other problem is likely to be the cause in this instance. Possible causes include: • Oxygen sensors have become wet • Current gas exposed to the Oxygen sensors is not what you think it is. • The mini-jack connection is corroded Check the readings on the STATUS screen (average) or Service Mode STATUS Screen (all 3 sensors). If a sensor is reading incorrectly, first remove it and clean the mini-jack connector. This can be done with Hollis approved electrical contact cleaner or simply by wiping the connector with a lint free cloth. DO NOT scratch the connector with a metal instrument. Then look for droplets of moisture on the sensor membrane. Use a rolled tissue to gently wick any moisture droplets from the sensor WITHOUT making contact with the sensor membrane itself. Pushing on the sensor membrane face can destroy the sensor. If neither of these corrects the reading, then the sensor should be replaced. The Oxygen sensors vary only slightly over time. Temperature, atmospheric pressure and moisture have far greater short-term effects on the readings. Calibrations carried out once a week should be more than sufficient, unless the EXPLORER has been transported to different climates or significantly different ambient conditions. | 41