Download E-625 Piezo Servo Controller
Transcript
6 Start-Up 6.3 Adjust the Sensor Zero-Point Changes in temperature or changes in the mechanical load can cause small deviations of the sensor zero point. Goal of the zero-point adjustment: Make the entire travel range available: When the sensor zero-point is set correctly, the complete output voltage range of the amplifier can be used in closed-loop operation. Prevent the piezo actuators from damage: In open-loop operation, the stage displacement with 0 V piezo voltage should already be about 10 % of the travel range (sensor monitor signal is +1 V). Then the average applied voltage is reduced which lengthens the lifetime of the piezo actuator in the stage without reducing the nominal travel range. INFORMATION The physical zero-position of the stage in closed-loop operation is changed by the zero-point adjustment of the sensor. Prerequisites You have read and understood the General Notes on Start-Up (p. 27). Only for the computer-controlled mode: − You have read and understood the user manual for the E-816 computer interface submodule. − The required software has been installed on the PC from the E-816 CD. You have installed the stage in the same manner as it will also be used in your application (corresponding load and alignment). The E-625 was installed (p. 19) properly. The E-625 is switched off, i.e. the wide-range-input power supply is not connected to the power socket over the power cord. Tools and Accessories 32 Only in analog mode: − Suitable analog signal source (p. 22). − SMB/BNC adapter cable (in scope of delivery) Version: 1.7.0 PZ166E E-625 Piezo Servo Controller