Download E-625 Piezo Servo Controller
Transcript
8 Adjustment of Internal Settings Synchronizing the Sensors Figure 10: Settings for the master/slave switch 1. Select an E-625 as the sensor master and check the position of the master/slave switch. The master position (see top figure) is the default setting. 2. If the master/slave switch on the selected sensor master is not in the master position, use the trimmer adjustment tool to turn it to the master position. 3. Set all other E-625 to sensor slave. To do this, turn the master/slave switches to the slave position (see top figure) using the trimmer adjustment tool. 4. Close the cases. 5. Make sure that all devices have the synchronization lines (pin 7 and 8 on the Network socket) and at least one ground wire (pin 1, 2 or 5 on the Network socket) connected to the corresponding wire, (for details see pin assignment of the E-625.CN network cable (p. 85)). E-625 Piezo Servo Controller PZ166E Version: 1.7.0 51