6 Start-Up
6.3 Adjust the Sensor Zero-Point
Changes in temperature or changes in the mechanical load can cause small
deviations of the sensor zero point.
Goal of the zero-point adjustment:

Make the entire travel range available:
When the sensor zero-point is set correctly, the complete output voltage range
of the amplifier can be used in closed-loop operation.

Prevent the piezo actuators from damage:
In open-loop operation, the stage displacement with 0 V piezo voltage should
already be about 10 % of the travel range (sensor monitor signal is +1 V).
Then the average applied voltage is reduced which lengthens the lifetime of
the piezo actuator in the stage without reducing the nominal travel range.
INFORMATION
The physical zero-position of the stage in closed-loop operation is changed by the
zero-point adjustment of the sensor.
Prerequisites

You have read and understood the General Notes on Start-Up (p. 27).

Only for the computer-controlled mode:
−
You have read and understood the user manual for the E-816 computer
interface submodule.
−
The required software has been installed on the PC from the E-816 CD.

You have installed the stage in the same manner as it will also be used in
your application (corresponding load and alignment).

The E-625 was installed (p. 19) properly.

The E-625 is switched off, i.e. the wide-range-input power supply is not
connected to the power socket over the power cord.
Tools and Accessories

32
Only in analog mode:
−
Suitable analog signal source (p. 22).
−
SMB/BNC adapter cable (in scope of delivery)