Function tables - summary of setup functions
Note: the order in which the functions appear on the display may not be exactly as shown below.
The availability and order of functions is determined by choice of function settings and options
fitted.
Functions in this first table are available in FUNC or CAL mode
Display
Function
Range
Default
Your Ref/Page
record
AxLo
Low setpoint value for
designated alarm relay x
Any display
value or OFF
OFF
See
4.1
5.1 / 13
AxHi
High setpoint value for
designated alarm relay x
Any display
value or OFF
OFF
See
4.1
5.2 / 13
AxHY
Hysteresis value for the
designated alarm relay x.
0 to 9999
10
See
4.1
5.3 / 14
Axtt
Trip time delay for the
designated alarm relay x.
0 to 9999
0
See
4.1
5.4 / 14
Axrt
Reset time delay for the
designated alarm relay x.
0 to 9999
0
See
4.1
5.5 / 15
Axn.o
or
Axn.c
Alarm relay x action to
normally open (de-energised) or
normally closed (energised)
Axn.o or
Axn.c
Axn.o
See
4.1
5.6 / 15
AxSP
or
Axt1
etc.
Relay operation independent
setpoint or trailing setpoint
(∗ Optional)
AxSP or
Axt1 etc.
AxSP
See
4.1
5.7 / 15
P.SEt
Preset value
Any display
value
0
5.8 / 16
brGt
Display brightness level
1 to 15
15
5.9 / 16
d.oFF
SECS
Auto display dimming timer
0 to 9999
0
5.10 / 16
(∗ Optional)—this function will only be accessible if the relevant option is fitted
Functions in this second table are available only in CAL mode or if ACCS is set to ALL
Display
Function
Range
Default
Your Ref/Page
record
REC_
Analog output option low
display value (∗ Optional)
Any display
value
0
5.11 / 17
REC~
Analog output option high
display value (∗ Optional)
Any display
value
1000
5.12 / 17
REC_
Ch2
Second analog output option
low display value (∗ Optional)
Any display
value
0
5.13 / 17
REC~
Ch2
Second analog output option
high display value (∗ Optional)
Any display
value
1000
5.14 / 18
(∗ Optional)—this function will only be accessible if the relevant option is fitted
8 of 33
RM4SSIMAN-2.2-0