Download HP E1445A Arbitrary Function Generator Module User`s Manual and

Transcript
[SOURce:]FREQuency2
[:CW | :FIXed]
[SOURce:]FREQuency2[:CW|:FIXed] <frequency> selects the sample rate for
arbitrary waveforms or the frequency for the standard waveforms (square, ramp,
triangle).
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<frequency>
numeric
see below | MINimum | MAXimum
Hz
Arbitrary Waveforms: MINimum selects the current reference oscillator frequency
divided by 131,072; MAXimum selects the current reference oscillator frequency.
Square Wave Outputs: MINimum selects the current reference oscillator frequency divided
by 524,288; MAXimum selects the current reference oscillator frequency divided by 4.
Ramps and Triangles Outputs: MINimum selects the current reference oscillator frequency
divided by 131,072 further divided by the SOURce:RAMP:POINts value; MAXimum selects
the current reference oscillator frequency divided by the SOURce:RAMP:POINts value.
The above values bound the valid range for <frequency>. The <frequency> value is
rounded to the nearest frequency that can be produced using the divide-by-n technique of
this generator.
Comments • If the actual frequency generated differs from the specified frequency by greater
than 1%, the Frequency bit of the Questionable Signal Status Register will be set.
See the STATus subsystem for more information.
• Executable when Initiated: Yes
• Coupling Group: Frequency
• Related Commands: TRIGger[:STARt]:SOURce, [SOURce:]FUNCtion[:SHAPe],
[SOURce:]ROSCillator commands, STATus subsystem
• *RST Condition: SOURce:FREQuency2:FIXed 10E3
Example Setting the Sample Rate or Waveform Frequency
FREQ2 1E3
Chapter 8
Sets frequency to 1000 Hz.
Command Reference
331