Download HERE - Nemoto
Transcript
SPECIFICATIONS: Output signal 600 +/- 150nA/ppm Cl2 NOTE: The output signal of the NE4-CL2 sensor is of opposite polarity to other sensors of this type – i.e. the signal is negative compared to (say) a standard CO or H2S sensor of the same type. Zero offset in clean air Response time (T90) Repeatability < +/- 0.2 ppm equivalent (T90) < 40 sec. (Same day measurement) < +/- 2% Zero offset < +/-0.2 ppm equivalent / year Sensitivity to CL2 < +/- 2% signal / month (Zero offset swing) < +/-0.5ppm (-20°C to +50°C) Long Term Stability Temperature dependence Minimum Detection Limit 0.1 ppm In Service -20°C to +50°C Recommended in Storage -0°C to +20°C In Service 15 – 90%RH (non-condensing) Recommended in Storage 15 – 90%RH (non-condensing) Temperature range Humidity range Pressure range in service 0.9 – 1.1 atm 0 – 10 ppm Detection range Maximum Overload Range Recommended load resistor 10Ω Recommended maximum storage time Warranty NE4-CL2-Manual, Issue 4, Sept 2015 50 ppm 6 months Against faulty workmanship or materials 24 months - page 4 of 15 -