Transcript
PA-4000 User Manual Ch3: Rev.1-1/05 3.1.5 Pre-Pulse Load Estimator The load is what the pulse voltage “sees” and it is a function of the material being electroporated. In a cuvette, this is a function of the cuvette plate spacing, the cuvette plate area and the conductivity of the aqueous solution. In tissue, it is a function of the electrode spacing and the conductivity of the tissue. The user may not know these. In order to estimate the load so an estimate of pulse amplitude can be provided, the PA-4000 uses a pre-pulse generated by the computer before the high voltage is turned on and the protocol is started. This pre pulse is 2.2 µs in duration and 2 volts in amplitude. The pulse is placed across the electrodes and the resulting current measured. Since the voltage is known, the resistance is calculated. This is done by the microprocessor and presented at the bottom of log report, as a resistance in ohms and a conductance in siemens, see Figure 3-1. The pre-pulse data is also used by the microprocessor to detect an output short or excessive conductivity of the ionic solution. If the load estimate is less than 10 ohms the high voltage will not be enabled and an “Output Shorted” message will appear. 3.1.6 Pulse Droop As explained in Section 2.2.3, rectangular wave electroporators also use storage or reservoir capacitors. When the high voltage is turned on this capacitor is charged to an initial voltage by the internal power supply. A pulse is generated when the capacitor is momentarily connected to the load and electrons in the capacitor are drained off like water running out of a reservoir. After a set time, the pulse width, the capacitor is disconnected from the load. Long pulses allow more electrons to run out and the voltage decreases (reservoir level drops) just as in an exponential discharge pulser. In rectangular wave electroporators the maximum pulse width is usually defined at the point that the pulse voltage at the end of the pulse is 95% of the initial voltage level. This voltage drop is called droop and is determined by the size of the internal reservoir capacitor and the load resistance. Droop is calculated by: Droop (%) = pulse width (sec onds ) C ∗ R load When using highly ionic loads, the electrons are depleted faster. Caution is required is setting pulse widths in these situations. Typical pulse widths yielding 5% droop or less in the PA-4000 are: 3-6 Load ohms 10 Low Range 4 to 400 volts High Range 5 to 1100 volts 420 µs 150 µs 20 300 µs 40 840 µs 1.68 ms 100 400 4.20 ms 20.0 ms 600 µs 1.50 ms 6.00 ms Cyto Pulse Sciences, Inc. P. O. Box 609, Columbia, MD 21045 410-787-1890