17-16 Calibration Model 6220/6221 Reference Manual Table 17-4 Remote current calibration summary Range Currents 100mA 100mA 0mA -100mA 20mA 0mA -20mA 2mA SOUR:CURR:RANG 2e-3 SOUR:CURR 2e-3 CAL:PROT:SOUR <+FS_DMM_Reading> SOUR:CURR 0 CAL:PROT:SOUR <0_DMM_Reading> SOUR:CURR -2e-3 CAL:PROT:SOUR <-FS_DMM_Reading> 2mA 0mA -2mA 200μA 200μA 0μA -200μA 20μA 0μA -20μA OUTP ON SOUR:CURR:RANG 0.1 SOUR:CURR 0.1 CAL:PROT:SOUR <+FS_DMM_Reading> SOUR:CURR 0 CAL:PROT:SOUR <0_DMM_Reading> SOUR:CURR -0.1 CAL:PROT:SOUR <-FS_DMM_Reading> SOUR:CURR:RANG 2e-2 SOUR:CURR 2e-2 CAL:PROT:SOUR <+FS_DMM_Reading> SOUR:CURR 0 CAL:PROT:SOUR <0_DMM_Reading> SOUR:CURR -2e-2 CAL:PROT:SOUR <-FS_DMM_Reading> 20mA 20μA Commands* SOUR:CURR:RANG 2e-4 SOUR:CURR 2e-4 CAL:PROT:SOUR <+FS_DMM_Reading> SOUR:CURR 0 CAL:PROT:SOUR <0_DMM_Reading> SOUR:CURR -2e-4 CAL:PROT:SOUR <-FS_DMM_Reading> SOUR:CURR:RANG 2e-5 SOUR:CURR 2e-5 CAL:PROT:SOUR <+FS_DMM_Reading> SOUR:CURR 0 CAL:PROT:SOUR <0_DMM_Reading> SOUR:CURR -2e-5 CAL:PROT:SOUR <-FS_DMM_Reading> * Use DMM (Figure 17-1) for 100mA-200nA ranges. * Use picoammeter (Figure 17-2) for 20nA and 2nA ranges. Return to Section 17 topics