Model 6220/6221 Reference Manual Calibration 17-19 Step 4: Calibrate guard circuit 1. Connect the digital multimeter to the Model 6220/6221 GUARD and LO jacks as shown in Figure 17-4. Be sure to connect the HI and LO terminals of the Model 6220/6221 OUTPUT jack together as shown. Select the DMM DC voltage function with auto range enabled. 2. Send these commands in order to calibrate the guard circuit using the <DMM_Reading> parameter (actual digital multimeter reading) for each step: SOUR:CURR:RANG 2e-4 SOUR:CURR 2e-5 OUTP ON CAL:PROT:GUAR:STEP1:INIT CAL:PROT:GUAR:STEP1 <DMM_Reading> CAL:PROT:GUAR:STEP2:INIT CAL:PROT:GUAR:STEP2 <DMM_Reading> OUTP OFF ‘ ‘ ‘ ‘ ‘ ‘ ‘ ‘ Set current range to 200μA. Set current source to 20μA. Turn on output. Set up for step 1 guard cal. Cal step 1 with reading. Set up for step 2 guard cal. Cal step 2 with reading. Turn off output. Step 5: Send dates, save calibration, and lock output calibration Send these command to send dates, save calibration constants and dates, and then finally lock out calibration: CAL:PROT:DATE 2004,05,15 CAL:PROT:NDUE 2005,05,15 CAL:PROT:SAVE CAL:PROT:LOCK ‘ ‘ ‘ ‘ Send cal date (5/15/2004). Send cal due date (5/15/2005). Save cal constants and dates. Lock out calibration. Return to Section 17 topics
Related documents
Model 6220 DC Current Source Model 6221 AC and DC Current
Christie CP2000-X User manual
Model 2182/2182A Nanovoltmeter
BodyMedia FIT Operating instructions
Manual - Test Equipment Depot
Christie CP2000 User Manual
Waveforms - Artisan Technology Group
Model 2520Pulsed Laser Diode Test System User`s Manual
AT-1 / AT-1smartprint ECG unit
AEMC Model 6550 Megohmmeter Manual PDF
Models 2500 And 2502 Dual Photodiode Meter User`s Manual
USER`S MANUAL