Download Alaris Medley Pump Module 8100 Version 4b

Transcript
Advisories
Advisory
Meaning
Response
Anesthesia Mode
Anesthesia Mode has been enabled.
None
Bolus Dose Complete
Channel is running in continuous
infusion mode if programmed.
None
Checking Line
A patient side occlusion has
occurred and AutoRestart feature is
monitoring downstream pressure to
determine if infusion can continue.
None
Infusion Complete - KVO
Programmed volume-to-be-infused
has been delivered; channel is
running at KVO rate.
Set up a new infusion or press
CHANNEL OFF key.
Panel Locked
Occurs following a key press when
Tamper Resist feature is active.
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of Programming
Module.
Panel Unlocked
Occurs when Tamper Resist feature is
deactivated.
None.
Pause
Pause control has been pressed;
infusion is stopped.
To resume infusion, press RESTART
key, or press CHANNEL SELECT key
and then press START soft key.
Secondary
A secondary infusion is in progress
on indicated channel.
None. When secondary VTBI=”0”,
infusion will revert to programmed
primary parameters.
48 ALARMS, ADVISORIES AND PROMPTS