R&S ESPI
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:LIMit<1...8>:TRACe
1 to 3
This command assigns a limit line to a trace.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
Example:
"CALC:LIM2:TRAC 3"
'Assigns limit line 2 to trace 3 in screen A.
"CALC2:LIM2:TRAC 1"
'Assigns limit line 2 to trace 1 in screen B at the same time.
Characteristics: *RST value: 1
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UNIT AMPere | DB | DBM | DBMV | DBPT | DBPW | DBUA | DBUA_M |
DBUV | DBUV_M | DBV | DEG | HZ | PCT | RAD | S | UNITLESS | VOLT | WATT
This command defines the unit for the associated limit line.
The definition is valid regardless of the measurement window.
Specification of the DB unit automatically switches the limit line over to relative mode. Units other than
DB cause the limit line to switch over to absolute mode.
Example:
"CALC:LIM4:UNIT DBUV"
'Sets the unit of limit line 4 to dBµV.
Characteristics: *RST value: DBM
SCPI: device-specific
Mode:
6.28
R, A
Operating Manual 1164.6607.12 - 02