R&S ESPI Remote Control – Description of Commands CALCulate Subsystem CALCulate<1|2>:LIMit<1...8>:TRACe 1 to 3 This command assigns a limit line to a trace. Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate the measurement window. Example: "CALC:LIM2:TRAC 3" 'Assigns limit line 2 to trace 3 in screen A. "CALC2:LIM2:TRAC 1" 'Assigns limit line 2 to trace 1 in screen B at the same time. Characteristics: *RST value: 1 SCPI: device-specific Mode: R, A CALCulate<1|2>:LIMit<1...8>:UNIT AMPere | DB | DBM | DBMV | DBPT | DBPW | DBUA | DBUA_M | DBUV | DBUV_M | DBV | DEG | HZ | PCT | RAD | S | UNITLESS | VOLT | WATT This command defines the unit for the associated limit line. The definition is valid regardless of the measurement window. Specification of the DB unit automatically switches the limit line over to relative mode. Units other than DB cause the limit line to switch over to absolute mode. Example: "CALC:LIM4:UNIT DBUV" 'Sets the unit of limit line 4 to dBµV. Characteristics: *RST value: DBM SCPI: device-specific Mode: 6.28 R, A Operating Manual 1164.6607.12 - 02