Download HP 4395A Service
Transcript
FIRMWARE INSTALLATION No rmware is installed in new A1 CPU assembly. When you replace a faulty A1 CPU with a new one, perform the following steps to install the rmware into the A1 CPU. Ordering the Firmware Diskette A rmware diskette (3.5 inch) that contains the analyzer's rmware is required for the rmware installation. If you do not have a rmware diskette, you must order one. For ordering information, contact your nearest Agilent Technologies service center and provide the revision of the analyzer's rmware. The part number of the rmware diskette depends on the rmware revision. The rmware revision of the analyzer is indicated on the revision label attached on the rear panel as shown in Figure 8-3. Figure 8-3. Firmware Revision Label Installing the Firmware Perform the following procedure to install the rmware into the analyzer. 1. Turn the analyzer power o. 2. Press both the 4Start5 and 4Preset5 keys. While pressing both keys, turn the analyzer power on. 3. Wait until the bootloader menu appears on the LCD as shown in Figure 8-4. 8-4 Digital Control Troubleshooting