Download HP 4395A Service
Transcript
FIND OUT WHY THE A1 +5 VD LED IS NOT ON STEADILY If the +5 VD LED is not on steadily, the +5 VD line voltage is missing or is not enough to power the analyzer. The problem may be in the A40 pre-regulator, the A50 DC-DC Converter, the A1 CPU, and any of assemblies obtaining the power from +5 VD supply. 1. Check the A40 Pre-Regulator a. b. c. d. Turn the analyzer power o. Disconnect a cable form the A50J1. The A50J1 location is shown in Figure 7-6. Turn the analyzer power on. Check the voltage between the pin 1 and pin 6(GND) of the cable within +22.0 V to +27.0 V using a voltmeter with a small probe. If the voltmeter reading is out of the limits, replace the A40 pre-regulator. If the voltmeter reading is within the limits, the A40 pre-regulator is veried. Turn the analyzer power o and reconnect the cable to the A50J1. Then continue with the next Check the A50 DC-DC Converter section. Figure 7-6. A40J1 Output Voltage 2. Check the A50 DC-DC Converter a. b. c. d. Turn the analyzer power o. Disconnect a cable form the A50J3. The A50J3 location is shown in Figure 7-6. Turn the analyzer power on. Check the voltage between the A50J3 pin 1 and pin 6(GND) within +4.59 V to +5.61 V using a voltmeter with a small probe. If the voltmeter reading is out of the limits, replace the A50 DC-DC Converter. Power Supply Troubleshooting 7-9