shown here to save complexity in the presentation. The electron
column is shown as a straight pipe. The restriction of pumping
speed by spray apertures and other features is represented as a
single aperture.
Two plotters are included, a dual pen plotter which shows the
pressures, respectively in the chamber and in the gun, and a
single pen plotter
which shows the
total oil mass
passed into the
chamber by the
diffusion pump.
This diffusion
pumped high
vacuum system
was replaced in
the mid 1970s by
a much simplified
turbo pumped
system, shown as
the second
example. Turbomolecular pump
prices, and their
reliability had
improved to the
point where an imaginative SEM manufacturer was able to take
the plunge. At that time the turbo pump was not accepted
generally as a reliable staple of high vacuum systems. The
move was a brave one.
The component count reduction can be seen by comparing the
pictorial schematics, and more dramatically by comparing the
VacSim schematics for the two systems.
This move radically reduced manufacturing costs and radically
improved system reliability. The change also imroved performance by reducing oily contamination in the SEM chamber and
51