TURNING ON HT CRT DISPLAY ● On VACUUM STATUS page check or wait until IGP < 23 and VACUUM STATUS: READY. Once IGP < 23, press READY button to return to main CRT page. ● On main CRT page select PARAMETERS page Check HIGH TENSION kV is set to 160 kV, and EMISSION is 1 or 2. (Higher EMISSION = brighter beam.) RIGHT HAND CONTROL PANEL ● Press HIGH TENSION ON/OFF to turn on HT (green light comes on). Wait until HT reaches 160 kV. ● On CRT page select FREE CONTROL. FOCUS on right hand control panel now controls HT. Turn large outer knob clockwise to gradually increase HT to 200 kV. If step size of HT increase is too small/large, turn inner knob STEP SIZE clockwise/ anticlockwise. Keep an eye on IGP and EMISSION; if either suddenly jump up, wait until they recover before further increasing HT. ● Once HT at 200 kV, press FREE CONTROL to deselect it, and press READY to return to main CRT page. 1 2 8