Download FLUOstar* OPTIMA Software User Manual
Transcript
FLUOstar1'OPTIMA
BMG Labtechnologies
Software User Manual
% CV of replicates
The CV (standard deviation / average) in percent of the replicates and the blanks will be
displayed.
Regression coeff. (r)
The regression coefficient as calculated from the linear regression equation is displayed.
Appears only if'Slope/secondCminute)1 is chosen in the calculation pull down menu.
Layout
The contents (standards, samples, blank) as defined in the layout section will be
displayed.
Standard concentration
The concentrations of the standards that were defined in the layout ('Concentrations and
Volumes' window) will be displayed.
Volumes 1, 2, 3, 4
The injection volumes for each pump, as defined in the layout ('Concentrations and
Volumes' window) will be displayed.
Sample IDs
Shows the Sample ID for each well.
Note:
If you use very long sample IDs you will only see a part here (approximately up to
10 characters). To see the full sample ID, use the Sample IDs worksheet.
Dilution factors
Shows the dilution factor for each well.
Multichromatic Data
If more than one filter setting was used in the test protocol (multichromatic) it is
possible to view the data for each setting by using the small pull-down box. The
box contains numbers corresponding to the order of the filter combinations. The
number is in the order in which the filters were defined in the test protocol (see
chapter 4.8 Multichromatics).
5
Choose the number in the box that corresponds to the raw data you want to view.
You can choose different numbers for each of the three tables to make
comparisons of the data.
Which filter setting correspondent to which position can be seen in the header of the
evaluation sheet. The pull-down box will be empty if only one filter pair has been used.
2002-11-15
0413F0003A
107/123