Step 5: Prepare the Micro-Centrifuge Tube Suspension
Items Needed


Inoculating loop containing organism cells
Micro-centrifuge tube from Step 3
Procedure
While holding the inoculating loop that contains the organism
cells:
1. Hold the water-filled micro-centrifuge tube in one
hand.
2. Open the tube and tilt it no more than 45º making
sure not to spill the contents.
3. Insert the inoculating loop into the water in the microcentrifuge tube and dislodge the sample into the
water by agitating the loop in the water as shown in Figure 24 – Dislodge the Organism
Cells into the Water in the MicroFigure 24.
centrifuge Tube
4. Withdraw the loop and immediately close the tube.
5. Properly dispose of the contaminated loop in a Biohazard Bag.
Step 6: Vortex the Sample
Items Needed


Micro-centrifuge tube with the organism cells in
dilution from step 5.
Vortex mixer
Procedure
1. Vortex the micro-centrifuge tube for at least 20
seconds, Figure 25.
2. Proceed directly to Step 7.
Note: To ensure optimum performance, do not delay
proceeding to Step 7.
Figure 25 – Vortex the Microcentrifuge Tube
MIT 1000 System User Guide V3.5