Assay 5 User Notes
Typical method parameters for method 5D are shown below. Values may vary
between instruments.
Method
Carrier Gas
Detector
Analysis Time (S)
Oven Temperature (°C)
Pressure (psi)
PID State
Injection Method
Inject Method
Inject Time (S)
Pump Time (S)
Backflush Time (S)
Integration Method
Automatic Integration
Slope Up (mV/S)
Slope Down (mV/S)
Min. Height (mV)
Min. Area (mV S)
Filter Level
Integration Delay (S)
Column B
Column C
Air
PID
600
60
12
High
Sensitivity
Air
PID
320
60
12
High
Sensitivity
Loop Injection
0.3
20
120
Loop Injection
0.8
20
20
Disabled
0.3
0.3
0.0
0
3
40
Disabled
1.0
1.0
0.0
0
3
20