Voyager User’s Guide
Method 5D
In method 5D, Voyager is configured for high concentrations of the sulfur gases.
Small sample volumes are injected. You must store the compounds at two or more
concentrations throughout the range to achieve maximum accuracy.
The table below lists the compounds and the typical concentration ranges for this
method. Concentrations outside the specified ranges, especially at the lower end, can
be obtained by adjusting the injection time, PID state and integration parameters.
Whenever accurate concentration readings are required, you should integrate peaks
manually and verify the integration.
Column B
Methyl Mercaptan
Ethyl Mercaptan
Dimethyl Sulfide
Dimethyl Disulfide
Column C
Hydrogen Sulfide
380
R.T. (S)
Concentration
Range (ppm)
Co-elution
80
98
103
545
0.5-50
0.5-50
0.5-50
0.5-50
No
Yes
Yes
No
32
0.3-100
No