Voyager User’s Guide Method 5D In method 5D, Voyager is configured for high concentrations of the sulfur gases. Small sample volumes are injected. You must store the compounds at two or more concentrations throughout the range to achieve maximum accuracy. The table below lists the compounds and the typical concentration ranges for this method. Concentrations outside the specified ranges, especially at the lower end, can be obtained by adjusting the injection time, PID state and integration parameters. Whenever accurate concentration readings are required, you should integrate peaks manually and verify the integration. Column B Methyl Mercaptan Ethyl Mercaptan Dimethyl Sulfide Dimethyl Disulfide Column C Hydrogen Sulfide 380 R.T. (S) Concentration Range (ppm) Co-elution 80 98 103 545 0.5-50 0.5-50 0.5-50 0.5-50 No Yes Yes No 32 0.3-100 No