Download AutovaxID Operations Manual

Transcript
Biovest International, Inc.
Primary User Interface Screens
Completed steps will be indicated with a checkmark, the current step will be
indicated in green, and steps that are incomplete or have been skipped will
be neither checked nor green.
NOTE: When running an instrument in “Strict Mode”, the user is not allowed to
deviate from the normal sequence of events as described above.
WARNING!
When using the Advanced Run Menu, the user should go through the stop
sequence in order to place the rotary valves in the load position for proper
removal of the cultureware. Failure to do so may split or damage the tubing of
the cultureware being removed or the next cultureware being loaded.
When a cell culture run is terminated normally using the Run Menu (see
Section 5.3), the AutovaxID positions the valves automatically.
AutovaxID Operations Manual
5-63