Download Manual - MOIST
Transcript
SECTION 6: Deploying the Instruments 6-3 Burst where the instrument continuously stores the selected Continuous mode burst parameter data at the selected sampling rate. Average where the instrument stores the selected current velocity Interval mode and tilt vector averages and one record of the selected instantaneous parameters at the end of every averaging interval for the Average On Time beginning at the start of every time interval. Burst where the 3D-ACM Wave continuously stores the selected Interval mode burst parameter data at the selected sampling rate for the Burst On Time beginning at the start of every time interval. Average and Burst Interleaved where the instrument operates in Interval mode Interval modeAverage for a specified number of time intervals, and then beginning at the start of the next time interval and for that time interval only, stores the selected burst parameter data at the selected sampling rate for the Burst On Time. This cycle repeats continuously. Average where the instrument commences the Interval Delayed mode modeAverage beginning at the Delayed Time on the Delayed Date. Burst where the instrument commences the Interval modeBurst, Delayed mode beginning at the Delayed Time on the Delayed Date. Average and Burst Interleaved where the instrument commences Delayed mode the Interval modeAverage and Burst Interleaved beginning at the Delayed Time on the Delayed Date. Note: For any Average operating mode, including Average and Burst Interleaved, if the Average Interval is less than 15 seconds, or if no current velocity or tilt vector averages are selected, or both, the selected instantaneous parameter data are stored at the selected sampling rate. Any selected vector averages are not stored. Note: For all Delayed running modes, if the Interval Time is zero, or if the Average On Time or burst On Time is greater than the Interval Time, then beginning at the Delayed Time on the Delayed Date the instrument commences Continuous modeAverage if Average is selected for the operating mode or Continuous modeBurst if Burst is selected for the operating mode. June 2003 3DACM97 Operating Manual