Download Solid State NMR: AVANCE Solids User Manual - Pascal-Man
Transcript
HETCOR with DUMBO, PMLG or w-PMLG, Using Shapes Table 8.1. Acquisition Parameters for pmlg-HETCOR (on tyrosine-HCl) Power level channel 2 for contact pulse. sp0 spnam0 ramp.100 or similar Shape for contact pulse channel f2. sp1 set to pl13 To match cnst20. spnam1 lgs-2 To match ppg calculation of in0, in_f1. p3 2.5 – 3 µsec 90º pulse channel 2 at pl12. p15 50 -500μs Contact pulse width. pcpd2 ≈ 2*p3 SPINAL64 /TPPM decoupling pulse. cpdprg2 SPINAL64/TPPM15 Decoupling sequence. l0 0 Start value 0, incremented during expt. l3 2- 4 Multiples of FSLG-periods, increment per row. in_f1 =in0 as calculated Set according to value calculated by ased. d1 2s Recycle delay. sw 310ppm Sweep width direct dimension. aq 16-20 msec masr 10-15 kHz F1 1H indirect F2 13C acquisition Table 8.2. 15 kHz ok at 100 kHz RF-field. Processing Parameters for pmlg-HETCOR (as for FSLG on tyrosine-HCl) Parameter Value Comment F2 direct dim 13C si 2-4 k wdw QSINE ph_mod pk SSB 2, 3 or 5. F1 indirect 1H si 256-1048 mc2 STATES-TPPI wdw QSINE ssb 3, 5 ph_mod pk User Manual Version 001 BRUKER BIOSPIN 87 (261)