Download User Manual - Accriva Diagnostics
Transcript
Section 2 Patient Sample Analysis OVERVIEW This section describes the procedure for performing a whole blood patient sample analysis on the IRMA TruPoint analyzer, including sample requirements, sample collection, and sample handling guidelines. SAMPLE REQUIREMENTS ACCEPTABLE SPECIMENS • Fresh arterial or venous whole blood collected in a 1, 2, or 3 mL lithium heparin syringe. Balanced or low-volume heparin is recommended for ionized calcium testing; sodium heparin may be used, but sodium values may be elevated 1-2 mmol/L. • Fresh capillary whole blood collected in the IRMA TruPoint Capillary Collection Device, which contains balanced lithium heparin. • Fresh venous whole blood collected in a lithium heparin collection tube. Balanced or low-volume heparin is recommended for ionized calcium testing; sodium heparin may be used, but sodium values may be elevated 1-2 mmol/L. The sample should be transferred to a non-heparinized 1, 2, or 3 mL syringe for injection into a cartridge. SYRINGE REQUIREMENTS Most standard ABG syringes are compatible with IRMA TruPoint cartridges. The following general types of syringes should not be used with IRMA TruPoint cartridges: • Frictionless or “pulsating” syringes. These syringes have plungers that will continue to travel downward after the user has stopped injecting. This may result in a sensor error. • Syringes that contain a mixing ball or non-dissolving disk impregnated with heparin. The ball or disk may become lodged in the tip of the syringe, and the sample may hemolyze when it is forced through or around the plug during injection. • Syringes that have a non-standard Luer hub that does not fit the IRMA TruPoint cartridge Luer injection port. 2.1