Download User Manual - Accriva Diagnostics
Transcript
Appendix F IRMA TruPoint Capillary Collection Device INTENDED USE The IRMA TruPoint Capillary Collection Device is intended for blood collection and the transfer of whole blood samples into IRMA TruPoint cartridges. DESCRIPTION Each capillary collection device contains 70 IU/mL balanced heparin and has a fill volume of 125uL. Each device consists of a capillary tube and a plunger, which is packaged separately. The capillary tube fills via capillary action until the sample contacts the white plug. The upper end of the capillary tube is sealed when blood contacts the white plug, preventing the sample from running back out the tip of the device during handling (see Figure F-1). Following sample collection, the plunger is inserted into the capillary tube, and the device is attached to the IRMA TruPoint cartridge Luer port. The sample is introduced into the cartridge by quickly and completely depressing the plunger in one smooth motion until it stops. Figure F-1 A. Capillary tube tip (round end where blood enters the capillary tube). B. Sample channel. C. Square flange. D. White plug. E. Capillary tube barrel (square end where plunger inserts). F.1