Download User Manual - Accriva Diagnostics
Transcript
a. Result outside of reference range: If reference ranges were established by the QA User, and a patient test result falls outside of the defined limits, the result will be flagged “H” (High) or “L” (Low). b. Result outside of reportable range: If a measured result is outside of the IRMA TruPoint reportable range, that result, and any corresponding calculated results, will be flagged “<” (less than) or “>” (greater than). c. Result suppressed: If a sensor errors during the analysis phase of a patient test, the result for that sensor, any sensors dependent on that sensor, and any corresponding calculated values, will be suppressed. Suppressed results will be dashed-out on both the screen and printout. Note: • pH Normalized iCa results will be reported only for pH values within the range of 7.2 - 7.6; if pH values fall outside of this range, pH Normalized iCa results will be suppressed. Complete Test Information Entry (optional); Document Patient Notes (optional) 23. Refer to Section 2-Test Information Entry and Section 2-Patient Notes for details. Obtain a Printout (optional) 24. a. If the Auto-print setting is “immediate”, test results will automatically print when the analysis is complete. b. If the Auto-print setting is “delayed”, test results will automatically print when the done button is pressed from the Results screen. c. If the Auto-print setting is “off”, a printout may be obtained by pressing the print button. 2.12