Download Model 5014i - User Manual
Transcript
Calibration Calibration Procedure 3. Cut/break the filter tape and remove from the bench. Then insert the Foil Holder from the calibration kit and press while maintaining a slight sideward pressure on the filter holder. MASS COEF: FOIL VALUE: 7000.0 1328 μg 3 REMOVE TAPE, INSERT FOIL HOLDER, TO CLOSE BENCH RANGE AVG DIAGS ALARM 4. Insert the Null/Zero foil with the label side up and press to proceed with the foil calibration, or press to stop the foil. MASS COEF:● FOIL VALUE: ● 7000.0 1328 µg MASS COEF: FOIL VALUE: Beta Avg 3 INSERT ●NULL FOIL TO START ZERO CAL AVG 3 PLEASE WAIT ● RANGE 7000.0 1328 µg 12016 1/sec DIAGS ALARM RANGE AVG 265 sec DIAGS ALARM The auto zero procedure of mass will begin and last approximately 270 seconds. When the zero is complete, the beta attenuation measurement will remain closed and an average Beta count rate is displayed. MASS COEF: FOIL VALUE: Beta Avg 7000.0 1328 μg 12022 1/sec 3 REMOVE NULL FOIL TO CONTINUE RANGE AVG DIAGS ALARM 5. Remove the zero/null foil and insert with the Span foil (e.g., 1328 ug) and press to begin the span calibration, or press to stop the foil. 4-10 Model 5014i Beta Instruction Manual Thermo Fisher Scientific