Download Model 5014i - User Manual
Transcript
Appendix C MODBUS Protocol This appendix provides a description of the MODBUS Protocol Interface and is supported both over RS-232/485 (RTU protocol) as well as TCP/IP over Ethernet. The MODBUS commands that are implemented are explained in detail in this document. The MODBUS protocol support for the iSeries enables the user to perform the functions of reading the various concentrations and other measurement values, read the status of the digital outputs of the analyzer, and to trigger or simulate the activation of a digital input to the instrument. This is achieved by using the supported MODBUS parameters listed below. Up to three simultaneous connections are supported over Ethernet. For details of the Model 5014i MODBUS Protocol specification, see the following topics: ● “Serial Communication Parameters” on page C-1 ● “TCP Communication Parameters” on page C-2 ● “Application Data Unit Definition” on page C-2 ● “Function Codes” on page C-3 ● “MODBUS Parameters Supported” on page C-8 Additional information on the MODBUS protocol can be obtained at http://www.modbus.org. References are from MODBUS Application Protocol Specification V1.1a MODBUS-IDA June 4, 2004. Serial Communication Parameters Thermo Fisher Scientific The following are the communication parameters that are used to configure the serial port of the iSeries to support MODBUS RTU protocol. Number of Data bits : 7 or 8 Number of Stop bits : 1 or 2 Parity : None, Odd, or Even Data rate : 1200 to 115200 Baud (9600 is default) Model 5014i Beta Instruction Manual C-1