Calibration
Calibration Procedure
3. Cut/break the filter tape and remove from the bench. Then insert the
Foil Holder from the calibration kit and press
while maintaining
a slight sideward pressure on the filter holder.
MASS COEF:
FOIL VALUE:
7000.0
1328 μg
3
REMOVE TAPE, INSERT FOIL
HOLDER, ƒ TO CLOSE BENCH
RANGE
AVG
DIAGS ALARM
4. Insert the Null/Zero foil with the label side up and press
to
proceed with the foil calibration, or press
to stop the foil.
MASS COEF:●
FOIL VALUE:
●
7000.0
1328 µg
MASS COEF:
FOIL VALUE:
Beta Avg
3
INSERT ●NULL FOIL
ƒ TO START ZERO CAL
AVG
3
PLEASE WAIT
●
RANGE
7000.0
1328 µg
12016 1/sec
DIAGS ALARM
RANGE
AVG
265 sec
DIAGS ALARM
The auto zero procedure of mass will begin and last approximately 270
seconds. When the zero is complete, the beta attenuation measurement
will remain closed and an average Beta count rate is displayed.
MASS COEF:
FOIL VALUE:
Beta Avg
7000.0
1328 μg
12022 1/sec
3
REMOVE NULL FOIL
ƒ TO CONTINUE
RANGE
AVG
DIAGS ALARM
5. Remove the zero/null foil and insert with the Span foil (e.g., 1328 ug)
and press
to begin the span calibration, or press
to stop
the foil.
4-10
Model 5014i Beta Instruction Manual