Servicing Detector Assembly Replacement Unhook Tubing from Measurement Head Unfasten Screws (4) Unhook Tubing from Rear Panel Unhook Connector from Measurement Interface Board “Beta 1” Figure 7–15. Replacing the Proportional Valve Detector Assembly Replacement Use the following procedure to replace the detector assembly (Figure 7– 16). Equipment Required: Detector assembly Hex driver, 4 mm Safety glasses WARNING The detector assembly should be replaced by a qualified technician knowledgeable in dealing with radiation precautions. By removing the detector assembly, the C-14 radioactive source is partially exposed. Safety glassed must be worn during this replacement procedure. The amount of C-14 meets the USNRC regulations as an exempt amount of radioactive source <100 Ci. ▲ Thermo Fisher Scientific Model 5014i Beta Instruction Manual 7-31