Download Agilent Technologies B1500A Technical data

Transcript
Remote Mode Functions
SPGU Module
Figure 2-31
Pulse Setup Parameters
SRP
V/10
Leading time
Pulse peak
V/10
V: Pulse level
Delay time
V/10
Pulse width
Pulseperiod
Td-sw
State hold time
Pulse switch
Normally open
Figure 2-32
Close
Trailing time
Pulse base
V/10
Delay time
Td-sw: Delay time for switching
Open
Close
Trigger Output in PG Mode
STGP 801,1
Pulse period/2 (maximum 5 s), TTL level
Trigger Output
0 V (chassis common) Pulse period
SPGU 801
base
Delay time
base
SPGU 802
base
SPUPD 801,901
SRP
Agilent B1500A/B1505A Programming Guide, Edition 11
base
SPP
2-57