Command Reference
BSM
2. The source channel forces the Stop value, and the monitor channel executes a
measurement.
If the search target value is out of the range between the measured value at the
Start value and the measured value at the Stop value, the search stops.
3. The source channel forces the Stop-D/2 value (or Stop+D/2 if Start>Stop), and
the monitor channel executes a measurement.
If the search stop condition is not satisfied, the measured data is used to decide
the direction (+ or –) of the next output change. The value of the change is
always half of the previous change.
4. Repeats the output change and measurement until the search stop condition is
satisfied.
For information on the search stop condition, see “BGI” or “BGV”. If the output
change value is less than the setting resolution, the search stops.
Figure 4-2
Binary Search Source Output Control Mode
Voltage or current
Stop
BSM, BST, and
BSV or BSI
-D/2
-D/8
-D/16
Search stopped
Delay time
(1) Normal output
Hold time
+D/32
+D/4
Trigger
+D/64
Delay time
Start
Stop
(2) Cautious output
BSM, BST, and
BSV or BSI
Delay time
-D/8
-D/16
Search stopped
Delay time
Hold time
+D/4
+D/32
+D/64
Trigger
Start
+D/2
Time
4-52
: Measurement
D = | Stop - Start |